111In Zevalin Imaging: Purpose

     To verify the expected biodistribution

     R/O rare cases of altered biodistribution

     Not to diagnose the extent of NHL

Data Acquisition for In-111 Zevalin

       Dual or single head gamma camera system

       Medium energy collimator(s)

       172 and 247 keV with 15% window

       256 x 1024 matrix

       Speed (dependent on time after injection)

      10 cm/min (20 minute scan) at 2 - 24 hours

        7 - 10 cm/min (30 minute scan) at 48 - 72 hours

•      5 - 7 cm/min (40 minute scan) at 90 - 120 hours (optional)

Expected Biodistribution

Anterior    Posterior

Anterior    Posterior

Anterior    Posterior

4-6 Hrs 64-70 Hrs 138-142 Hrs

Representative Patient:  111In-Labeled Zevalin® Monoclonal Antibody Imaging

4 hour 66 hours 139 hours

Abdominal SPECT

Abdominal CT
 

Altered Biodistribution Image

 

Median Y-90 Zevalin Radiation
Absorbed Doses to Organs and Tumor