Murphy’s
Law favors the Inspector!
-
Linearity checks are usually performed
on a timely basis, but often are not taken out to the lowest dose level one
would use clinically. A reasonable level is 30 mCi, comparable to doses of
iodinated compounds or Cr-51 labeled RBCs
-
Inspectors often see people drawing
doses, injecting patients, eluting generators and handling eluate without
gloves. Be prepared during an inspection- wear gloves if there is a chance
you will be working with radioactivity in the inspector’s presence.
-
Survey meters must undergo a brief
operational check before each use. Results must be recorded at least
quarterly as well as after calibration, repair, or battery change.
-
Room survey records are not complete
unless they include the make, model, and serial # of the instrument used. In
addition, if a well counter is used for wipe tests, the make, model and
serial # must be recorded in the logbook.
-
When performing the Constancy Test on
dose calibrators, one MUST check every setting that might be used that day.
For example, when called in for a stat lung scan, both the Tc-99m and Xe-133
settings should be checked. Often a ventilation study is added on and one
forgets to check the calibrator for Xe-133.
-
An
annual review of the radiation safety program must be performed by either
the Radiation Safety Officer or his designate (may be a consultant). Results
must be reviewed by Hospital Administration.
-
An
annual review of the ALARA Program must also take place.
-
Brachytherapy records must indicate
complete accountability for records of all inventory items. There should be
a conversion factor for converting Radium units to mCi.
-
Rooms in which Xe-133 gas are used must
be under negative pressure. Inspectors find failures occasionally when
construction is underway and airflow is diverted from Nuclear Medicine to
another location, destroying the required pressure differential.
-
Frequent failure for unit dose users is
the calibration, dating, timing, and initialing the dose record.
-
A minor issue is Continuing Education
credits. Most NMTs are conscientious, but occasionally license renewal is
held up due to failure to obtain enough credits.
-
All authorized users are required to
undergo radiation safety training on an annual basis.
-
Hand monitoring might be required
daily, or every time you leave the laboratory, depending upon license
requirements for your institution.
-
Inspectors like to review High Dose
Iodine therapy procedures, especially issues related to dose calibration,
patient dose administration procedures and safety precautions taken, and
waste disposal records.
-
Incoming package logbooks are kept up
well, in general. Inspectors are most likely to ask to see log-in of a
package received on the weekend or on a holiday.
-
Regarding administration of I-131 Na
iodide: If the patient has been admitted, then, regardless of the dose of
I-131 (whether 8 mCi or 200 mCi), nursing instructions must be distributed,
signage must be posted on the door, room surveys must be performed, and all
required precautions for the hospitalized radioactive patient must be
observed.
-
Regarding generator QC testing, the
most common failure is to not report Mo-breakthrough as a ratio, e.g., 0.01
mCi mCi/mCi Tc. Those people still using generators are very conscientious
about performing the test, even when called in for a stat scan.
-
NRC has proposed a check of
Mo-breakthrough and Al ion breakthrough at time of first elution. If tests
pass, no additional testing is required.
-
NOT APPLICABLE IN AGREEMENT STATES
